A simple microwave kit to convert an X-band ESR spectrometer for K-band operation

Bhagat, Vinod R. ; Venkataraman, Balu (1977) A simple microwave kit to convert an X-band ESR spectrometer for K-band operation Pramana - Journal of Physics, 9 (6). pp. 605-612. ISSN 0304-4289

[img]
Preview
PDF - Publisher Version
756kB

Official URL: http://www.ias.ac.in/j_archive/pramana/9/6/605-612...

Related URL: http://dx.doi.org/10.1007/BF02848147

Abstract

A kit for the conversion of an X-bandesr spectrometer for K-band operation is described and its performance has been tested with p-benzosemiquinone, vanadyl acetyl acetonate and 2, 5-ditertiary butyl semiquinone. The easy conversion of the X-band spectrometer to K-band operation enables us to distinguish between magnetic field dependent and field independent parameters. A varactor harmonic generator is used as a frequency doubler driving the fundamental power from the existing X-band source to give a power output at K-band without having to modify the electronic circuit; the modification of the microwave circuitry is proposed.

Item Type:Article
Source:Copyright of this article belongs to Indian Academy of Sciences.
Keywords:ESR; K-band Instrumentation; X-band Spectrometer; Microwave Circuitry; Varactor Harmonic Generator