Evaluation and comparison of pulsed and continuous wave radiofrequency electron paramagnetic resonance techniques for in vivo detection and imaging of free radicals

Yamada, Ken-Ichi ; Murugesan, Ramachandran ; Devasahayam, Nallathamby ; Cook, John A. ; Mitchell, James B. ; Subramanian, Sankaran ; Krishna, Murali C. (2002) Evaluation and comparison of pulsed and continuous wave radiofrequency electron paramagnetic resonance techniques for in vivo detection and imaging of free radicals Journal of Magnetic Resonance, 154 (2). pp. 287-297. ISSN 0022-2364

Full text not available from this repository.

Official URL: http://www.sciencedirect.com/science/article/pii/S...

Related URL: http://dx.doi.org/10.1006/jmre.2001.2487

Abstract

The performance of two electron paramagnetic resonance (EPR) spectrometers/imagers, one configured in pulsed mode and the other in continuous wave (CW) mode, at an operating frequency of 300 MHz is compared. Using the same resonator (except for altered Q-factors), identical samples and filling factors in the two techniques have been evaluated for their potentials and limitations for in vivo spectroscopic and imaging applications. The assessment is based on metrics such as sensitivity, spatial and temporal resolution, field of view, image artifacts, viable spin probes, and subjects of study. The spectrometer dead time limits the pulsed technique to samples with long phase memories (>275 ns). Nevertheless, for viable narrow-line spin probes, the pulsed technique offers better sensitivity and temporal resolution. The CW technique, on the other hand, does not restrict the choice at spin probes. In addition, the phase-sensitive narrow-band detection of the CW technique gives artifact-free images even for large objects. Selected examples illustrating the performance of the CW and pulsed techniques are presented to put the capabilities of the two techniques in perspective.

Item Type:Article
Source:Copyright of this article belongs to Elsevier Science.
ID Code:51927
Deposited On:01 Aug 2011 07:35
Last Modified:01 Aug 2011 07:35

Repository Staff Only: item control page