Continuous shipboard sampling system for determination of triple oxygen isotopes and O2/Ar ratio by dual-inlet mass spectrometry

Sarma, V. V. S. S. ; Abe, O. ; Yoshida, N. ; Saino, T. (2006) Continuous shipboard sampling system for determination of triple oxygen isotopes and O2/Ar ratio by dual-inlet mass spectrometry Rapid Communications in Mass Spectrometry, 20 (23). pp. 3503-3508. ISSN 0951-4198

Full text not available from this repository.

Official URL: http://onlinelibrary.wiley.com/doi/10.1002/rcm.275...

Related URL: http://dx.doi.org/10.1002/rcm.2756

Abstract

A continuous shipboard sampling system was developed for the determination of the isotopic composition of the triple oxygen isotopes and oxygen to argon (O2/Ar) ratios in dissolved air. In this system, dissolved air is separated by a hollow fiber membrane degassing module. This system collects dissolved air quantitatively and rapidly. The sample flow rate through the membrane is critical for the fractionation of the oxygen isotopes and the O2/Ar ratio and should be <2 mL/min. Fractionation of oxygen between the liquid and gas phase of the air-saturated water was found to be similar to that of earlier reports. The advantages of this method over existing techniques include rapid collection of samples (30 min/sample), high efficiency in extraction of gases from the liquid phase, and the lack of a sample preparation step (e.g. degassing).

Item Type:Article
Source:Copyright of this article belongs to John Wiley and Sons, Inc.
ID Code:105737
Deposited On:01 Feb 2018 12:16
Last Modified:01 Feb 2018 12:16