Calibration of the CMS drift tube chambers and measurement of the drift velocity with cosmic rays

Gurtu, A. ; et., al (2010) Calibration of the CMS drift tube chambers and measurement of the drift velocity with cosmic rays Journal of Instrumentation, 5 (3). T03016_1-T03016_38. ISSN 1748-0221

Full text not available from this repository.

Official URL: http://iopscience.iop.org/1748-0221/5/03/T03016/

Related URL: http://dx.doi.org/10.1088/1748-0221/5/03/T03016

Abstract

This paper describes the calibration procedure for the drift tubes of the CMS barrel muon system and reports the main results obtained with data collected during a high statistics cosmic ray data-taking period. The main goal of the calibration is to determine, for each drift cell, the minimum time delay for signals relative to the trigger, accounting for the drift velocity within the cell. The accuracy of the calibration procedure is influenced by the random arrival time of the cosmic muons relative to the LHC clock cycle. A more refined analysis of the drift velocity was performed during the offline reconstruction phase, which takes into account this feature of cosmic ray events.

Item Type:Article
Source:Copyright of this article belongs to Institute of Physics.
Keywords:Particle Tracking Detectors (gaseous Detectors); Large Detector Systems for Particle and Astroparticle Physics
ID Code:92218
Deposited On:29 May 2012 12:11
Last Modified:29 May 2012 12:11

Repository Staff Only: item control page