An automated thermal relaxation calorimeter for operation at low temperature (0.5K<T<10K)

Banerjee, S. ; Prins, M. W. J. ; Rajeev, K. P. ; Raychaudhuri, A. K. (1992) An automated thermal relaxation calorimeter for operation at low temperature (0.5K<T<10K) Pramana - Journal of Physics, 39 (4). pp. 391-404. ISSN 0304-4289

[img]
Preview
PDF - Publisher Version
935kB

Official URL: http://www.ias.ac.in/j_archive/pramana/39/4/391-40...

Related URL: http://dx.doi.org/10.1007/BF02845823

Abstract

We describe an automated calorimeter for measurement of specific heat in the temperature range 10 K>T>0.5 K. It uses sample of moderate size (100-1000 mg), has a moderate precision and accuracy (2%-5%), is easy to operate and the measurements can be done quickly with He4 economy. The accuracy of this calorimeter was checked by measurement of specific heat of copper and that of aluminium near its superconducting transition temperature.

Item Type:Article
Source:Copyright of this article belongs to Indian Academy of Sciences.
Keywords:Relaxation Calorimeter; Specific Heat; Low Temperature